CCD Spectroradiometer Integrating Sphere Compact System | 燈飾 | 照明配件及零件
CCD Spectroradiometer Integrating Sphere Compact System
中國內地
|
生產線所在地:中國內地
展位編號: GH-P09 | 測檢及驗証服務 |
產品規格
產品簡介

LPCE-3 is a CCD Spectroradiometer Integrating Sphere Compact System for LED Testing. It is suitable for single LED and LED luminaires’ photometric, colorimetric and electrical measurement. The measured data meets the requirements of CIE 177CIE84,  CIE-13.3COMMISSION DELEGATED REGULATION (EU) 2019/2015IES LM-79-19Optical-Engineering-49-3-033602IESNA LM-63-2ANSI-C78.377 and GB standards.

Standard:
LM-79-19 “Optical and Electrical Measurements of Solid State Lighting Products”
IESNA LM-63-2 “IESNA Standard File Format for the Electronic Transfer of Photometric Data and Related information”
CIE 177 “Color Rendering of White LED Light Sources”
CIE84 “The measurement of luminous flux”
CIE-13.3 “Method of Measuring and Specifying Colour Rendering Properties of Light Sources”
ANSI-C78.377 “Specifications for the Chromaticity of Solid State Lighting products for Electric Lamps”
(EU) 2019/2015 “COMMISSION DELEGATED REGULATION”

Configuration: 
CCD Spectroradiometer (LMS-7000), Optical Fiber (CFO-1.5M), Digital Power Meter (LS2012), DC Power Source (DC-S Series), AC Power Source (LSP-500VAS), Integrating Sphere (IS-1.5MA-CASE and IS-0.3M), Standard Light Source (SLS-50W and SLS-10W)

Measurement:
• Colorimetric: Chromaticity coordinates, CCT, Color Ratio, Peak Wavelength, Half Bandwidth, Dominant Wavelength, Color Purity, CRI (Ra, R1 to R15), Spectrum Test, TM30 (Rf, Rg), CQS
• Photometric: Luminous Flux, Luminous Efficiency, Radiant Power, EEI, PAR, PPF
• Electrical: Voltage, Current, Power, Power Factor, Harmonic (Optional)

Specification:
• Spectral Range Wavelength: 350nm~950nm(LMS-7000VIS) (200~1050nm is optional)
• Spectral Wavelength Accuracy: ±0.5nm Wavelength Reproducibility: ±0.5nm
• Sample Scanning Steps: ±0.1nm
• Accuracy of Chromaticity Coordinate (Δx, Δy): ±0.003 (under Standard A Lamp)
• Correlated Color Temperature CCT: 1,500K~25,000K. CCT Accuracy: ±0.5%
• Color Rendering Index Range: 0~100.0, Accuracy: ± (0.3%rd±0.3)
• Photometric linear: ±0.5%
• Stray light: <0.015% (600nm) and <0.03% (435nm)
• The system includes an auxiliary lamp device and the software includes the self-absorption coefficient correction function
• Connect with PC via USB cable. The English version software can run in Win7, Win8, Win10, Win11 (Microsoft certified driver)

生產所需時間

20